1. The operation and daily maintenance of the pressure steam sterilizer should be carried out by well-trained personnel.

2. Preventive maintenance procedures should include regular inspection of the sterilizer cabinet cavity, door tightness, and all instruments and controls by qualified personnel.

3. Saturated steam should be used, and it does not contain corrosive inhibitors or other chemicals that may contaminate the items being sterilized.

4. All items to be autoclaved should be placed in a container with good air permeability and good heat permeability; the sterilizer cabinet should be loosely loaded so that the steam can act on the load evenly.

pressure steam sterilizer

5. When the interior of the sterilizer is pressurized, the interlocking safety device can prevent the door from being opened, while the pressure steam sterilizer without interlocking device should close the main steam valve and open the door when the temperature drops below 80°C.

6. When the liquid is autoclaved, since it may boil due to overheating when the liquid is taken out, the slow discharge setting should be used.

7. Even if the temperature of the pressure steam sterilizer drops below 80°C, the operator should wear appropriate gloves and face shield for protection when opening the door.

Keyword:

Scan the QR code to read on your phone

Relevant Information

The use of Dengguan steam sterilizer

The use of Dengguan steam sterilizer

The company's steam sterilizer is a safe and reliable sterilization equipment that adopts automatic program control, which is mainly composed of a shell, a sterilization chamber, a steam generator, a piping system, and a control system.
Boosting Infection Control Efficiency with Table Top Steam Sterilizers

Boosting Infection Control Efficiency with Table Top Steam Sterilizers

Discover how table top steam sterilizers can revolutionize infection control protocols in healthcare facilities, ensuring a safer and more hygienic environment for patients and staff alike.
The Importance of Anti Decubitus Mattress in the Healthcare Industry

The Importance of Anti Decubitus Mattress in the Healthcare Industry

Discover the significance of anti decubitus mattresses in the medical equipment sector. Learn about how these specialized mattresses help prevent pressure ulcers and provide comfort to patients in nee
上一页
1
2
...
187

CONTACT US

Jiangsu Dengguan Medical Equipment Co., Ltd.

Tel:0086-519-82421222

Address: No. 17, Danfeng West Road, Jintan District, Changzhou City, Jiangsu Province
SEO Tags

Webpage © Jiangsu Dengguan Medical Equipment Co., Ltd.  All Rights Reserved.  苏ICP备2021032787号     powered by: www.300.cn

Jiangsu Dengguan Medical Equipment Co., Ltd.

苏ICP备2021032787号

powered by: www.300.cn